Maharani, Shafira Putri and Firdy Yuana, S.Si., M.Si and Chomsin Sulistya Widodo, S.Si., M.Si., Ph.D. (2024) Estimasi Entrance Surface Air Kerma Dan Analisis Diagnostic Reference Level Berdasarkan Pengukuran Tube Output Pada Pemeriksaan Thorax Radiografi Umum. Sarjana thesis, Universitas Brawijaya.
Abstract
Penggunaan pesawat radiografi umum secara berulang dapat menurunkan kualitas alat dan berpotensi membahayakan kesehatan pasien. DRL (Diagnostic Reference Level) merupakan salah satu cara untuk mengoptimalkan kerja pesawat radiografi umum. Penelitian ini bertujuan untuk mengetahui nilai ESAK (Entrance Surface Air Kerma) dan menghitung DRL (Diagnostic Reference Level) untuk ESAK dan DAP (Dose Air Product) pada pemeriksaan thorax dengan radiografi umum di Rumah Sakit Budhi Asih, Jakarta. Penelitian ini menggunakan data 681 pasien yang menjalani pemeriksaan thorax dengan radiografi umum dari Agustus hingga Desember 2023. Nilai ESAK dihitung berdasarkan pengukuran Tube Output, dan nilai DRL dihitung berdasarkan nilai ESAK dan DAP. Hasilnya dibandingkan dengan iDRL (Indonesian Diagnostic Reference Level) dan DRL Internasional. Nilai DRL untuk ESAK dan DAP berturut-turut adalah 0,13 mGy dan 0,013 Gy.cm2 . Dari nilai tersebut, dapat disimpulkan bahwa pesawat radiografi umum, di unit radiologi Rumah Sakit Umum Daerah Budhi Asih, masih dalam batas aman dan tidak melebihi iDRL maupun DRL internasional.
English Abstract
The repeated use of general radiography can impair its quality and harm the patient's health. DRL (Diagnostic Reference Level) is one method for optimizing the performance of general radiography. The purpose of this study is to determine the ESAK (Entrance Surface Air Kerma) value and calculate the DRL (Diagnostic Reference Level) for ESAK and DAP (Dose Air Product) during a thorax examination using general radiography at Budhi Asih Hospital in Jakarta. This study examined data from 681 participants who had a thoracic examination with general radiography between August and December 2023. The ESAK value was obtained using Tube Output measurements, whereas the DRL was determined using the ESAK and DAP values. The results were compared to the Indonesian Diagnostic Reference Level (iDRL) and the International DRL. The DRL for ESAK and DAP are 0.13 mGy and 0.013 Gy.cm2 , respectively. These figures indicate that the general radiography equipment in the radiology unit at Budhi Asih Regional General Hospital remains within safe limits and does not exceed the iDRL or the International DRL.
Item Type: | Thesis (Sarjana) |
---|---|
Identification Number: | 052409 |
Uncontrolled Keywords: | Radiografi Umum, Tube Output, ESAK, DAP, DRL |
Divisions: | Fakultas Matematika dan Ilmu Pengetahuan Alam > Fisika |
Depositing User: | Unnamed user with username nova |
Date Deposited: | 01 Oct 2024 01:38 |
Last Modified: | 01 Oct 2024 01:38 |
URI: | http://repository.ub.ac.id/id/eprint/225448 |
Text (DALAM MASA EMBARGO)
Shafira Putri Maharani.pdf Restricted to Registered users only Download (2MB) |
Actions (login required)
View Item |