Djatmika, Rosalina (2014) Pengaruh pH dan Ion Asing Terhadap Kinerja Elektroda Selektif Ion CdCl3- Tipe Kawat Terlapis Berbasis Alliquat 336-CdCl3- Untuk Penentuan Kadar Kadmium dalam Air Limbah Baterai. Sarjana thesis, Universitas Brawijaya.
Abstract
Kadmium merupakan logam berat pencemar lingkungan perairan yang bersifat toksik bagi makhluk hidup. Ion kadmium dalam perairan dapat ditentukan melalui metode potensiometri menggunakan elektroda selektif ion (ESI) CdCl3- tipe kawat terlapis bermembran dengan komposisi perbandingan (% berat) Aliquat 336-CdCl3- : PVC : DBP = 0,04 : 0,33 : 0,63 dalam pelarut tetrahidrofuran (THF) dengan perbandingan b/v = 1 : 3 yang telah dikarakterisasi kinerjanya yang didasarkan pada harga Faktor Nernst. Tujuan penelitian ini adalah mempelajari pengaruh pH dan ion asing terhadap kinerja ESI CdCl3- tipe kawat terlapis untuk menentukan kadar kadmim dalam air limbah baterai. Pengaruh pH ditentukan dengan mengukur potensial larutan CdCl3- pada pH 1-7. Pengaruh ion asing ditentukan menggunakan metode larutan tercampur dengan konsentrasi ion asing NO3- dan H2PO4- masing-masing 0,01 M dalam larutan ion utama CdCl3- dalam rentang konsentrasi 10-1-10-6 M. ESI CdCl3- tipe kawat terlapis selanjutnya diaplikasikan untuk menentukan kadar kadmium dalam air limbah baterai. Hasil penelitian menunjukkan bahwa ESI CdCl3- tipe kawat terlapis dapat bekerja optimal pada pH 1-4 dan tidak terganggu dengan kehadiran ion asing NO3- dan H2PO4-. Urutan selektifitas ESI CdCl3- tipe kawat terlapis terhadap ion asing adalah :CdCl3->NO3->H2PO4- . ESI CdCl3- digunakan untuk menentukan kadar ion kadmium pada air limbah baterai, diperoleh kadar kadmium sebesar 13,6 ± 0,34ppm ion CdCl3- yang memberikan akurasi sebesar 98,98 %.
English Abstract
Cadmium is a heavy metal pollutants in marine environment that are toxic to living beings. Cadmium ions in water can be determined by potentiometric method by using coated wire CdCl3- ion selective electrode (CdCl3- ion CWE) which has a membrane with a composition ratio (% by weight) of Alliquat-336-CDCl3-: PVC: DBP=0.04: 0.33: 0.63 dissolved in tetrahydrofuran (THF) solvent with a ratio of 1:3 (w/v) which has characterized its performance based on he Nernst factor prices.The purpose of this study was to determine the influence of pH and foreign ions on the performance of CdCl3-ion CWE for determination of cadmium levels in battery waste water. The influence of pH was determined by measuring the potential of CdCl3-ion solution in a concentration range of 10-1-10-6 M at pH 1-7. Effect of foreign ions was determined using the mixed solution method with foreign ions NO3- and H2PO4- concentration of 0.01 M respectively and CdCl3- primary ion in a concentration range of 10-1-10-6 M. The results showed that coated wire CdCl3 -ion selective electrode (CdCl3-ion CWE) can work optimally at pH 1-4 and was not disturbed by the presence of foreign ions NO3-and H2PO4- The sequence selectivity of CdCl3-ion CWE against foreign ion is CdCl3->NO3-> H2PO4-. ISE CdCl3- was used to determined cadmium level in battery waste water, which contain 13,6 ± 0,34 ppm CdCl3 -ion has an accuracy of 98,98%
Item Type: | Thesis (Sarjana) |
---|---|
Identification Number: | SKR/MIPA/2014/60/051400905 |
Subjects: | 500 Natural sciences and mathematics > 540 Chemistry and allied sciences |
Divisions: | Fakultas Matematika dan Ilmu Pengetahuan Alam > Kimia |
Depositing User: | Budi Wahyono Wahyono |
Date Deposited: | 11 Feb 2014 09:26 |
Last Modified: | 11 Feb 2014 09:26 |
URI: | http://repository.ub.ac.id/id/eprint/154075 |
Actions (login required)
![]() |
View Item |