Validasi Metode Analisis Barium (Ba) Dalam Air Limbah Menggunakan Inductively Coupled Plasma – Optical Emission Spectrometry (ICP-OES) Berdasarkan SNI 6989-82:2018

Helda, - and Prof. Akhmad Sabarudin, S.Si., M.Sc., Dr.Sc and Seldy Marcelino, A.Md.Si (2025) Validasi Metode Analisis Barium (Ba) Dalam Air Limbah Menggunakan Inductively Coupled Plasma – Optical Emission Spectrometry (ICP-OES) Berdasarkan SNI 6989-82:2018. Sarjana thesis, Universitas Brawijaya.

Abstract

Metode baku yang dimodifikasi harus dilakukan validasi untuk memastikan bahwa metode mampu menghasilkan data yang akurat, presisi, dan andal. Penelitian ini bertujuan untuk melakukan validasi metode analisis logam barium dalam air limbah menggunakan Inductively Coupled Plasma-Optical Emission Spectroscopy (ICPOES). Parameter pengujian validasi meliputi linieritas, presisi keterulangan, akurasi, limit deteksi metode (LDM), limit kuantisasi (LK), robustness, dan estimasi ketidakpastian pengukuran. Berdasarkan hasil penelitian, diperoleh nilai koefisien korelasi (r) sebesar 0,99998, %SBR sebesar 0,44%, %recovery berada dalam rentang 87,93% hingga 88,90%, nilai LDM dan LK masing-masing sebesar 0,0055 mg/L dan 0,0200 mg/L, serta nilai Fhitung dan thitung masing-masing sebesar 1,5672 dan 0,2066. Hasil pengujian memenuhi syarat keberteriman yang telah ditetapkan dengan nilai ketidakpastian sebesar 0,1828±0,0065 mg/L. Dengan demikian, metode telah terbukti valid dan dapat diandalkan untuk analisis rutin kadar barium dalam sampel air limbah di Laboratorium Uji PT MMID

English Abstract

The modified standard method must be validated to ensure that it provides accurate, precise, and reliable results. This study aims to validate the analysis method for barium in wastewater using Inductively Coupled Plasma-Optical Emission Spectroscopy (ICP-OES). The validation parameters include linearity, repeatability precision, accuracy, method detection limit (MDL), limit of quantification (LoQ), robustness, and measurement uncertainty. The test results show a correlation coefficient (r) of 0.99998, a %RSD of 0.44%, %recovery ranging from 87.93% to 88.90%, MDL and LoQ values of 0.0055 mg/L and 0.0200 mg/L, respectively, and Fcalculated and tcalculated values of 1.5672 and 0.2066, respectively. The test results meet the acceptance criteria with an uncertainty value of 0.1828 ± 0.0065 mg/L. Therefore, the method is confirmed to be reliable and valid for routine analysis of barium concentration in wastewater samples at the PT MMID Testing Laboratory

Item Type: Thesis (Sarjana)
Identification Number: 052509
Uncontrolled Keywords: logam, barium, validasi metode, ketidakpastian pengukuran, ICP-OES
Subjects: 500 Natural sciences and mathematics > 540 Chemistry and allied sciences
Divisions: Fakultas Sains, Teknologi dan Matematika > Kimia
Depositing User: Unnamed user with email admin
Date Deposited: 05 Mar 2026 05:58
Last Modified: 05 Mar 2026 05:58
URI: http://repository.ub.ac.id/id/eprint/243921
[thumbnail of DALAM MASA EMBARGO] Text (DALAM MASA EMBARGO)
HELDA.pdf
Restricted to Registered users only

Download (6MB)

Actions (login required)

View Item View Item