Analisa Optimisasi Dan Limitasi Dosis Dalam Pemeriksaan Foto Thorax Dengan Pesawat Mobile X Ray Di Ruang IGD

Ariady, Achmad Farid (2019) Analisa Optimisasi Dan Limitasi Dosis Dalam Pemeriksaan Foto Thorax Dengan Pesawat Mobile X Ray Di Ruang IGD. Sarjana thesis, Universitas Brawijaya.

Abstract

Pemanfaatan sinar-X di ruang IGD seringkali dilakukan di luar ruang radiologi. Namun hal tersebut memerlukan perhatian yang khusus atas efek negatif yang dapat ditimbulkan oleh radiasi sinar-X baik itu terhadap para petugas IGD, pekerja radiasi, pasien dan masyarakat di sekitar yang berpotensi terpapar radiasi. Atas dasar tersebut, telah dilakukan pengukuran laju dosis paparan radiasi dan menganalisa jarak aman, serta perhitungan besar dosis yang diterima pasien selama pemeriksaan foto thorax di ruang P2 IGD dengan menggunakan pesawat sinar-X mobile. Proses pengukuran dan analisa ini didahului oleh uji kesesuaian guna mengetahui kondisi pesawat yang akan digunakan untuk penelitian. Selanjutnya jika hasil pengujian telah terlewati dilanjutkan dengan pengukuran laju dosis paparan radiasi di sekitar sumber dan menghitung dosis yang diterima pasien. Dari hasil pengukuran tersebut kemudian dianalisa tentang kesesuaian laju dosis paparan radiasi pada jarak yang bervariasi dengan nilai batas dosis (NBD) yang ditetapkan Bapeten. Tujuan penelitian ini adalah untuk mencari berapa besar dosis paparan radiasi dari pesawat sinar-X mobile yang berpotensi membahayakan bagi tubuh manusia serta mengetahui dosis yang diterima pasien pada pemeriksaan foto thorax, yang kemudian akan diketahui jarak aman dari paparan radiasi baik itu dengan menggunakan proteksi radiasi maupun tanpa proteksi radiasi. Manfaat penelitian ini adalah sebagai salah satu upaya untuk meningkatkan proteksi dan keselamatan radiasi terutama dalam prinsip optimasi dan limitasi dosis.

English Abstract

The use of X-rays in the emergency room is often done outside the radiology room. However, this requires special attention to the negative effects that can be caused by X-ray radiation both on emergency staff, radiation workers, patients and the surrounding community who are potentially exposed to radiation. On this basis, a measurement of the dose rate of radiation exposure and analyzing the safe distance have been carried out, as well as the calculation of the dose received by the patient during chest X-ray examination in the emergency room specially in P2 room, using a mobile X-ray. This measurement and analysis process is preceded by a suitability test to determine the condition of the mobile X-ray to be used for research. Furthermore, if the test results have passed, then measure the dose rate of radiation exposure around the source and calculate the dose received by the patient. The results of the measurements were then analyzed about the suitability of the radiation exposure dose rate at varying distances with the dose limit value (NBD) set by Bapeten. The purpose of this study was to find out how many doses were potentially harmful to the human body and find out the dose received by patients on chest x-ray examination, which would then be known to be safe from radiation exposure using radiation protection or without radiation protection. The benefits of this research are as an effort to increase radiation protection and safety especially in the principle of optimization and limitation of doses.

Other obstract

-

Item Type: Thesis (Sarjana)
Identification Number: SKR/MIPA/2019/344/052001499
Uncontrolled Keywords: Paparan dosis, Optimasi, Limitasi, NBD. Exposure to Dosage, Optimization, Limitation, DRL.
Subjects: 500 Natural sciences and mathematics > 519 Probabilities and applied mathematics > 519.6 Mathematical optimization
Divisions: Fakultas Matematika dan Ilmu Pengetahuan Alam > Fisika
Depositing User: Budi Wahyono Wahyono
Date Deposited: 10 Aug 2020 07:46
Last Modified: 10 Aug 2020 07:46
URI: http://repository.ub.ac.id/id/eprint/179815