Abstract

Elektroda Selektif Ion (ESI) Hg2+ tipe kawat terlapis dibuat menggunakan bahan aktif kitosan, dengan polivinilklorida (PVC) sebagai bahan pendukung, dioktilphtalat (DOP) sebagai bahan pemlastis, dengan perbandingan kitosan : PVC : DOP = 3 : 39 : 58 dalam pelarut tetrahidrofuran (THF) (1:3 b/v) dengan waktu perendaman optimum selama 40 menit. ESI Hg(II) tipe kawat terlapis berbasis kitosan menunjukkan harga faktor Nernst -29,8 mV/dekade konsentrasi, rentang konsentrasi linier 10-4 – 10-1 M, limit deteksi 9,77x10-5 M atau setara 26,53 ppm, dengan waktu respon 200 detik. Penelitian ini bertujuan untuk mengaplikasikan metode ESI Hg(II) tipe kawat terlapis berbasis kitosan pada pengukuran kadar merkuri dalam sampel sediaan kosmetik. Hasilnya dibandingkan dengan metode standar spektrofotometri sinar tampak menggunakan uji t. Berdasarkan hasil uji t, pengukuran menggunakan metode potensiometri dengan ESI Hg2+ tipe kawat terlapis berbasis kitosan tidak berbeda secara bermakna ditunjukkan dari nilai thitung < ttabel. Hasil validasi metode dan dihasilkan akurasi dan presisi yang baik yaitu sebesar 98,67% dan 94,33%. Oleh karena itu, metode ESI Hg2+ tipe kawat terlapis berbasis kitosan dapat digunakan sebagai metode alternatif untuk pengukuran kadar merkuri dalam sampel sediaan kosmetik.

English Abstract

A coated wire mercury ion selective electrode (Hg(II)-ISE) were prepared using chitosan as ionophore, supported by a mixture of polyvinylchloride (PVC) as polymer, dioctylpthalate (DOP) as plasticizer with ratio (% weight) of chitosan: PVC: DOP = 3: 39: 58 dissolved in tetrahydrofuran (THF) (1:3 w/v) solvent with 40 minutes of soaking time. This Hg(II)-ISE showed Nernst factor of -29,8 mV/decade of Hg(II) concentration under linier concentration range from 1x10-4 – 1x10-1 M, detection limit of measurement of 9,77x10-5 M or 26,53 ppm, and response time of 200 seconds. The purpose of this study is to apply the Hg2+ - ISE on the determination of Hg2+ in cosmetic samples. The results obtained using Hg2+ - ISE were compared to those obtained by standard method visible spectrophotometry with reagent of dithizone. Based on the result of statistical t test, the measurement of Hg2+ ion using potensiometry method did not differ significantly with those obtained from visible spectrophotometry method with accuracy and precision values of 98,67% and 94,33%. Therefore, Hg2+ - ISE can be used as an alternative method for the measurement of Hg2+ level in cosmetic samples.

Other obstract

-

Item Type: Thesis (Sarjana)
Identification Number: SKR/FK/2014/506/051403777
Uncontrolled Keywords: merkuri, potensiometri, ESI, kitosan, kosmetik, validasi; mercury, potentiometric, ISE, chitosan, cosmetics, validation
Subjects: 600 Technology (Applied sciences) > 615 Pharmacology and therapeutics > 615.9 Toxicology > 615.925 Specific inorganic poisons
Divisions: Fakultas Kedokteran > Farmasi
Depositing User: Hasbi
Date Deposited: 11 Jul 2014 09:40
Last Modified: 03 Dec 2021 07:43
URI: http://repository.ub.ac.id/id/eprint/124508
[thumbnail of SKRIPSI_IRWINDA GRAFIYAN PUTRA.pdf]
Preview
Text
SKRIPSI_IRWINDA GRAFIYAN PUTRA.pdf

Download (4MB) | Preview

Actions (login required)

View Item View Item